Flow cell leaks on ACQUITY TUV detector - WKB287996

Article number: 287996

SYMPTOMS

  • Leaks collect under the detector 
  • No peaks 

ENVIRONMENT

  • ACQUITY UPLC Tunable UV Detector

CAUSE

Broken flow cell 

FIX or WORKAROUND

  1. Replace the flow cell (205015016 - ACQUITY UPLC ID Cell TUV, Analytical) 

ADDITIONAL INFORMATION

  • ACQUITY Cell TUV, Standard (205015000)
  • ACQUITY UPLC TUV Flow Cell, Low Flow (205015001)
  • ACQUITY UPLC ID Cell TUV, HS FC (205015018)
  • ACQUITY UPLC TUV Flow cell, 10 mm 500nL (205015033)
  • nanoTUV Flow Cell (205015013)
  • Flow Cell Shunt (700002825)
  • ACQUITY UPLC TUV 5 mm SS FC (205000610)
  • ACQUITY UPLC TUV 5 mm Titanium FC (205000611)
  • Refurb ACQ TUV Flowcell, Analytical 10mm (800000231)
  • Refurb ACQ TUV Flowcell, High Sens, 25mm (800000232)

  •  

Not able to find a solution? Click here to request help.