1515 pump leaks at the inlet manifold - WKB28677

Article number: 28677

SYMPTOMS

  • 1515 pump leaks at the inlet manifold
  • Reseating and tightening the tubing does not resolve the issue

ENVIRONMENT

  • 1515 Isocratic HPLC Pump

CAUSE

Worn or damaged fittings

FIX or WORKAROUND

Replace the solvent inlet tubing (700001486).

ADDITIONAL INFORMATION