Baseline drift seen with 2998 Photodiode Array Detector - WKB283118

Article number: 283118

SYMPTOMS

  • Downward baseline drift seen with 2998 Photodiode Array Detector
  • Flow cell cleaning performed
  • Lamp hours <500 
  • Lamp energy approx. 15k at 230 nm 

ENVIRONMENT

  • 2998 Photodiode Array Detector

CAUSE

Low lamp energy

FIX or WORKAROUND

Replace the lamp (201000281).

ADDITIONAL INFORMATION