High pressure from ACQUITY Premier - WKB281466
Article number: 281466
SYMPTOMS
- System pressure is very high with very low flow (0.1 mL/min)
- Pressure is normal with no column
ENVIRONMENT
- ACQUITY Premier
- ACQUITY Arc Premier
CAUSE
Clogged column
FIX or WORKAROUND
Flush the column appropriately, or replace the column if necessary.
ADDITIONAL INFORMATION
If pressure remains high with no column, clean the system as described in article wkb116495, without a column attached.