How to disassemble Acquity UPLC Column Manager - WKB280365

Article number: 280365

OBJECTIVE or GOAL

Shut down and disassemble an ACQUITY Column Manager for shipment to the repair center.

ENVIRONMENT

  • ACQUITY UPLC Column Manager

PROCEDURE

  1. Wash the system with H2O for 20 minutes, and then wash the system with IPA for 20 minutes.
  2. Close the control software and power-off all modules on the stack.
  3. Electrically unplug all UPLC modules and mark and remove any cables (Ethernet, signal, I/O, events) from the modules.
    (Marking the cables is important for reinstalling the system when the modules come back from repair. Photos of the cable and tubing are also helpful.)

    clipboard_e32c3c6f3700cd53bfb6df95e8dc0e4f3.png
     
  4. Remove the fluidics connections:

           clipboard_e390b814d7cd3fc254342458ce95e7364.png                    
         
  5. Move the module to a suitable box for shipment. (If a box is not available, ask your local Waters office.)
    Note the dimensions and weight of the module to choose the right box:
    ACQUITY UPLC Column Manager: 34.4 cm x 61.0 cm x 20.0 cm, 21.0 kg
  6. Fill and sign the latest revision of the H&S module provided by Waters.
  7. Attach it to the box and send a PDF copy to the local Waters office.

  8. The Waters office will contact you to arrange the shipment.