Low pressure with ACQUITY Sample Manager - WKB277887

Article number: 277887

SYMPTOMS

  • Error received when attempting to prime wash syringes: "Sample pressure low [value]; possibly no wash solvent or needle damaged"
  • Solvent pooling under Sample Manager
  • No solvent puddles in sample compartment; sample needle is not bent
  • Priming wash solvent does not clear error

ENVIRONMENT

  • ACQUITY UPC2 System
  • ACQUITY UPLC I-Class

CAUSE

Broken and/or leaking wash syringe

FIX or WORKAROUND

Replace the wash syringe. Part 700006070 (SYRINGE, 2.5 mL, INVERTED) is used for both UPC2 and I-Class SM-FL.

ADDITIONAL INFORMATION

 

id277887, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL