Low pressure with ACQUITY Sample Manager - WKB277887
Article number: 277887
SYMPTOMS
- Error received when attempting to prime wash syringes: "Sample pressure low [value]; possibly no wash solvent or needle damaged"
- Solvent pooling under Sample Manager
- No solvent puddles in sample compartment; sample needle is not bent
- Priming wash solvent does not clear error
ENVIRONMENT
- ACQUITY UPC2 System
- ACQUITY UPLC I-Class
CAUSE
Broken and/or leaking wash syringe
FIX or WORKAROUND
Replace the wash syringe. Part 700006070 (SYRINGE, 2.5 mL, INVERTED) is used for both UPC2 and I-Class SM-FL.
ADDITIONAL INFORMATION
id277887, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL