Skip to main content
Waters

High baseline noise and random spikes in the chromatography - WKB276278

Article number: 276278

SYMPTOMS

  • No pressure rippling 
  • Detector flow cell is clean 
  • Baseline stable with zero injection volume 
  • Replacing the in-line filter does not solve the problem 

ENVIRONMENT

  • 2695 Alliance system
  • e2695 Alliance system
  • 2489 TUV detector

CAUSE

The Inlet tubing on the injector HPMV1 is worn or partially clogged.

FIX or WORKAROUND

Replace the inlet tubing on the injector HPMV1.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.