When the manual Rheodyne injector on the 1515 Isocratic HPLC Pump is switched from Load to Inject, no data is acquired, no Auto Zero - WKB27625
Article number: 27625
SYMPTOMS
- Software reads "Waiting for Injection" after an injection is made and the injector is switched from Load to Inject
ENVIRONMENT
- 1515 Isocratic HPLC Pump
- Breeze 2
- Rheodyne manual inject valve
CAUSE
The inject start cable is not connected properly.
FIX or WORKAROUND
Ensure that the inject start I/O connection on the back of the 1515 pump (slots 1 and 2) is connected to the inject start I/O connection on the back of the detector.
If auto zero is desired connect a second cable from inject start on 1525 to auto zero on the detector