Skip to main content
Waters

Low system pressure on ACQUITY Premier UPLC system - WKB274300

Article number: 274300

SYMPTOMS

  • Low system pressure on ACQUITY Premier UPLC system
  • Solvent manager passes dynamic leak test on both pumps
  • Needle seal readiness test failure

ENVIRONMENT

  • Empower
  • ACQUITY Premier

CAUSE

The sample needle is damaged.

FIX or WORKAROUND

Replace the sample needle (700012820).

ADDITIONAL INFORMATION

 

id274300, DLT

Not able to find a solution? Click here to request help.