Article number: 274300

SYMPTOMS

  • Low system pressure on ACQUITY Premier UPLC system
  • Solvent manager passes dynamic leak test on both pumps
  • Needle seal readiness test failure

ENVIRONMENT

  • Empower
  • ACQUITY Premier

CAUSE

The sample needle is damaged.

FIX or WORKAROUND

Replace the sample needle (700012820).

clipboard_86ea08a1-b1a3-4d99-8ec5-18c85d9b05a8.png

ADDITIONAL INFORMATION