Arc HPLC pressure is very high when using flow path 2 - WKB274103

Article number: 274103

SYMPTOMS

  • Arc HPLC pressure is very high when using flow path 2
  • Pressure is normal when using flow path 1

ENVIRONMENT

  • Arc HPLC

CAUSE

The mixer for flow path 2 has a blockage.

FIX or WORKAROUND

Replace the mixer.

ADDITIONAL INFORMATION

 