Arc HPLC pressure is very high when using flow path 2 - WKB274103
Article number: 274103
SYMPTOMS
- Arc HPLC pressure is very high when using flow path 2
- Pressure is normal when using flow path 1
ENVIRONMENT
- Arc HPLC
CAUSE
The mixer for flow path 2 has a blockage.
FIX or WORKAROUND
Replace the mixer.
ADDITIONAL INFORMATION