Instrument Failure: 2475 serial number uninitialized - WKB2696

Article number: 2696

SYMPTOMS

  • Instrument Failure when trying to run 2475 in Breeze
  • Battery recently replaced and serial number reset in detector

ENVIRONMENT

  • Breeze 2

CAUSE

Node configuration made when the 2475 did not have a serial number specified on the detector.

FIX or WORKAROUND

  1. Delete system in Breeze.
  2. Recreate system in Breeze using the 2475 with the serial number specified.
  3. Delete any instance of a 2475 serial number uninitialized.

ADDITIONAL INFORMATION