Instrument Failure in Empower or Breeze: Serial number uninitialized in Node Properties - WKB2696

Article number: 2696

SYMPTOMS

  • Instrument Failure when trying to run system
  • Battery is new
  • Node Properties shows Serial number uninitialized for instrument status, status = no
  • Duplicate entry of instrument found in Node Properties with correct serial number and status = yes
  • "Not Connected" within Run Samples window of Empower

ENVIRONMENT

  • Breeze 2
  • Empower 3
  • Waters Ethernet modules

CAUSE

Node Properties does not specify instrument serial number.

FIX or WORKAROUND

  1. In node properties, delete the line that says status = no,  Serial number uninitialized
  2. Click on Scan Instruments
  3. Ensure status of the instrument modules say status = Yes
  4. Recreate system in software
  5. Launch Run Samples to test 