Instrument Failure in Empower or Breeze: Serial number uninitialized in Node Properties - WKB2696
Article number: 2696
SYMPTOMS
- Instrument Failure when trying to run system
- Battery is new
- Node Properties shows Serial number uninitialized for instrument status, status = no
- Duplicate entry of instrument found in Node Properties with correct serial number and status = yes
- "Not Connected" within Run Samples window of Empower
ENVIRONMENT
- Breeze 2
- Empower 3
- Waters Ethernet modules
CAUSE
Node Properties does not specify instrument serial number.
FIX or WORKAROUND
- In node properties, delete the line that says status = no, Serial number uninitialized
- Click on Scan Instruments
- Ensure status of the instrument modules say status = Yes
- Recreate system in software
- Launch Run Samples to test
ADDITIONAL INFORMATION