UPC2 ABPR not reaching setpoint - WKB267368

Article number: 267368

SYMPTOMS

  • ABPR not reaching setpoint when running
  • ABPR health check passing

ENVIRONMENT

  • UPC2

CAUSE

Column fitting leaking

 

FIX or WORKAROUND

  1. Re-attach column fitting as per KCS: WKB52762

ADDITIONAL INFORMATION

 