Repeated failure of turbo molecular pump on an ACQUITY QDa Detector - WKB26724
Article number: 26724
SYMPTOMS
- Repeated failure of turbo molecular pump on an ACQUITY QDa (requiring replacement pumps).
ENVIRONMENT
- ACQUITY QDa
- Turbo molecular pump
- Detector control card (DCC)
CAUSE
A faulty detector control card (DCC) (700011090) at the "Turbo Pump Controller" at junction 6 (J6), which supplies the 24V power supply to the turbo molecular pump. This led to over power supply, causing premature failure of the turbo pump controller.
FIX or WORKAROUND
Replace the DCC board (700011090).
ADDITIONAL INFORMATION
id26724, QDA