Error flag: Sample manager FTN: temperature controller watchdog time-out - WKB26651

Article number: 26651

SYMPTOMS

  • Error flag occurs in middle of sequence: "Sample manager FTN: temperature controller watchdog time-out"
  • Chiller drifts in and out of preset alarm band range
  • Ambient temperature in lab is close to 24 °C and FTN struggles to cool

ENVIRONMENT

  • FTN H-Class, H-Class Bio, I-Class
  • Empower 2, Empower 3

CAUSE

Failed sample chiller

FIX or WORKAROUND

Replace the sample chiller unit (700011357).

ADDITIONAL INFORMATION