nanoACQUITY error "Pump A: Pump HW over pressure" - WKB26611

Article number: 26611

SYMPTOMS

  • Error "Pump A: Pump HW over pressure" appears as soon you switch from trapping mode to analytical
  • Removing the Analytical column, Trapping column, and Sample Manger out of the flow path doesn't help
  • Removing the Flow Controller doesn't help; after reconnecting, the error reappears

ENVIRONMENT

  • nanoACQUITY UPLC
  • ACQUITY UPLC M-Class

CAUSE

 

FIX or WORKAROUND

  1. Connect Magic Mix with 1% FA and flush the system for 12 hours.

  2. Flush back system to first 50:50 MeOH/H2O for 12 hours.

  3. Flush back to customer solvents.

ADDITIONAL INFORMATION

 

 

id26611, methanol, UPMASM, UPMBSM, UPMSMFL, UPNAYXPMP, UPNSAMMGR, UPNSOLMGR