nanoACQUITY error "Pump A: Pump HW over pressure" - WKB26611
Article number: 26611
SYMPTOMS
- Error "Pump A: Pump HW over pressure" appears as soon you switch from trapping mode to analytical
- Removing the Analytical column, Trapping column, and Sample Manger out of the flow path doesn't help
- Removing the Flow Controller doesn't help; after reconnecting, the error reappears
ENVIRONMENT
- nanoACQUITY UPLC
- ACQUITY UPLC M-Class
CAUSE
FIX or WORKAROUND
-
Connect Magic Mix with 1% FA and flush the system for 12 hours.
-
Flush back system to first 50:50 MeOH/H2O for 12 hours.
-
Flush back to customer solvents.
ADDITIONAL INFORMATION