FTN does not start - WKB263471

Article number: 263471

SYMPTOMS

  • FTN does not start
  • Reinstalling the firmware does not help
  • When measuring the voltage of the CPU2000 battery, it is a little low, 3.0 V, so that is not the cause

ENVIRONMENT

  • FTN

CAUSE

Bad CPU2000

FIX or WORKAROUND

Replace the CCPU, 3V (700013063).

ADDITIONAL INFORMATION

WAT080443 - Battery, 3V Lithium