Compression check failing on Alliance 2695 Separations Module (high pressure valve needs to be rebuilt) - WKB2627

Article number: 2627

SYMPTOMS

  • Compression check fails
  • New seal pack
  • New syringe
  • Purge injector several times
  • Solvent leaking on valve housing
  • Restriction missing

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 717plus Autosampler

CAUSE

V1 and/or V2 valve (high-pressure valve) needs to be rebuilt

FIX or WORKAROUND

Rebuild V1 valve and/or V2 valve (High Pressure Motorized Valve Rebuild Kit [WAT045424])