How to calibrate R axis - WKB261247
Article number: 261247
OBJECTIVE or GOAL
Calibrate R axis.
ENVIRONMENT
- ACQUITY UPLC H-Class Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY Premier Sample Manager - Flow Through Needle (SM-FTN)
- Arc/Arc HPLC Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY UPLC I-Class Sample Manager - Flow Through Needle
- ACQUITY UPLC I-Class Sample Manager - Fixed Loop
- ACQUITY Premier Sample Manager - Fixed Loop
- ACQUITY UPC2 Sample Manager
PROCEDURE
- Enter service mode in the ACQUITY Console. On the troubleshooting menu, select "Service Mode" and type the password.
- In the Sample Manager section of the Console, choose the Maintain pull-down menu, select "Calibrate R-axis", and follow the Wizard prompts.
- Remove plates and click Next.
- Insert the short side of the calibration pin (part 405013532; usually located in the foam insulation in the Sample Manager door) in the calibration hole and click Next.
Calibration pin has a small target dimple in the top.
- Use the +Z button to drive the needle down to within 2-3 mm of the calibration pin hole on the top. Switch the displacement increment to 0.1 mm or 1 mm as required.
- Switch the displacement increment to 0.1 mm and move the needle along the R axis (+R or -R) to align the tip of the needle and the calibration pin hole.
- When the tip of the needle and the Calibration hole are aligned, press "Calibrate", and then press "Save".
- A warning message appears as a reminder to remove the calibration pin. Press OK. The calibration pin can be removed after the Sample Manager is reset.
- From the troubleshooting menu, disable service mode.