"Needle/Seal move limit exceeded (5.0)" error on ACQUITY UPLC H-Class FTN - WKB260440
Article number: 260440
SYMPTOMS
- Error "Needle/Seal move limit exceeded (5.0)"
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class
CAUSE
Bent, broken or dislodged needle
FIX or WORKAROUND
- Inspect the needle and if dislodged, re-install the needle correctly. If the needle is installed properly and the error persists continue with step 2.
- Replace the needle.
- Calibrate R axis and needle Z axis.
- Characterize the needle seal (seal constant value should be between 8 and 11 N/mm).
- Perform the Needle Seal Readiness test.