ACQUITY UPLC HTCH Ethernet switch: Failure to connect instrument modules to Waters DHCP server - WKB2603
Article number: 2603
SYMPTOMS
-
System not able to connect or communicate through the network card or Ethernet switch PCB.
-
No lights are visible at any point of operation on the Ethernet switch PCB.
-
Instrument modules are not visible in Waters DHCP Server and Loader Utility when going through Ethernet switch PCB.
-
Modules successfully visible in Waters DHCP Server and Loader Utility when bypassing the Ethernet switch PCB and connecting directly to each module from PC.
ENVIRONMENT
- ACQUITY UPLC Sample Manager.
- ACQUITY UPLC High Temperature Column Heater (HTCH).
CAUSE
Failed external control cable. No power to Ethernet switch PCB.
FIX or WORKAROUND
Replace part number 441000871 for the CABLE, MOLDED, D-SUB, M-F, 2.5 FT LONG (External Control Cable from ACQUITY UPLC Sample Manager to HTCH).
ADDITIONAL INFORMATION
id2603, UPBINARY