Article number: 259781

SYMPTOMS

  • PDA baseline is not stable

ENVIRONMENT

  • ACQUITY Arc

CAUSE

Contaminated or clogged flow cell

FIX or WORKAROUND

  1. Remove the column and connect to the detector with a union.
  2. Flush all the mobile phase lines with 100% water at a flow rate of 0.2 mL/min for 3 hours.
  3. Flush all the mobile phase lines with 100% IPA at a flow rate of 0.2 mL/min for 3 hours.
  4. Flush all the mobile phase lines with 100% water at a flow rate of 0.2 mL/min for 1 hour.
  5. Flush all the mobile phase lines with 1% formic acid in water at a flow rate of 0.2 mL/min for 3 hours.
  6. Flush all the mobile phase lines with 2% ammonia in water at a flow rate of 0.2 mL/min for 3 hours.
  7. Flush all the mobile phase lines with 100% water at a flow rate of 0.2 mL/min for 3 hours.
  8. Use wet prime when replacing the solvent.

ADDITIONAL INFORMATION

 

id259781, A-30PDA, eluent, isopropanol, UPAPDA, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA