Baseline is not stable on ACQUITY Arc PDA /TUV - WKB259781
Article number: 259781
SYMPTOMS
- PDA baseline is not stable
- TUV baseline is not stable
ENVIRONMENT
- ACQUITY Arc
CAUSE
Contaminated or clogged flow cell
FIX or WORKAROUND
- Remove the column and connect to the detector with a union.
- Flush all the mobile phase lines with 100% water at a flow rate of 0.2 mL/min for 3 hours.
- Flush all the mobile phase lines with 100% IPA at a flow rate of 0.2 mL/min for 3 hours.
- Flush all the mobile phase lines with 100% water at a flow rate of 0.2 mL/min for 1 hour.
- Flush all the mobile phase lines with 1% formic acid in water at a flow rate of 0.2 mL/min for 3 hours.
- Flush all the mobile phase lines with 2% ammonia in water at a flow rate of 0.2 mL/min for 3 hours.
- Flush all the mobile phase lines with 100% water at a flow rate of 0.2 mL/min for 3 hours.
- Use wet prime when replacing the solvent.
ADDITIONAL INFORMATION