Skip to main content
Waters

Alliance compression check failure with 200-µL sample loop installed - WKB25932

Article number: 25932

SYMPTOMS

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Compressible volume is increased when conducting the compression test with larger sample loops.

  • The parameters for the compression test were much too tight at anything below 20-psi delta decay with the larger sample loop.

FIX or WORKAROUND

  1. Adjust the compression test parameters according to the default parameters recommended in the operators guide.
    • Recommendation: use a max delta decay of 50 psi as the default.
    • Change other parameters, such as the target compression pressure, only as necessary when working with larger sample loops and/or different solvent compressibility.
  2. Review the compression test guidelines outlined in the following operator's guides:

Compression Check Diagnostic Parameters.jpg

ADDITIONAL INFORMATION

Compression test passed with repeatable results with the 200-µL sample loop when delta decay was 50 psi and target compression pressure was 200 µL. This is within the acceptable range of parameters for characterizing the instrument using this diagnostic.

Not able to find a solution? Click here to request help.