How to disassemble a 2475 HPLC FLR Detector for shipment- WKB257792
Article number: 257792
OBJECTIVE or GOAL
Shut down and disassemble a 2475 HPLC FLR Detector for shipment to the repair center.
ENVIRONMENT
- 2475 FLR HPLC Detector
- 2475 FLR Arc/Arc HPLC Detector
PROCEDURE
- Wash the system with H2O for 20 minutes, and then wash the system with IPA for 20 minutes.
- Close the control software and power-off the module.
- Electrically unplug the module and mark and remove any cables (Ethernet, signal, I/O, events) from the module.
(Marking the cables is important for reinstalling the system when the modules come back from repair. Photos of cable and tubing are also helpful.)
- Disconnect inlet and outlet tubing from the flow cell.
- Move the module to a suitable box for shipment. (If a box is not available, ask your local Waters office.)
Note the dimensions and weight of the module to choose the appropriate box:
• 2475 Blue Detector: 20.8 cm X 34.3 cm X 61.0 cm, 17.2 kg
• 2475 Platinum Detector: 20.8 cm X 28.4 cm X 50.3 cm, 13.6 kg - Fill and sign the latest revision of the H&S module provided by Waters.
-
Attach it to the box and send a PDF copy to the local Waters office.
-
The Waters office will contact you to arrange the shipment.
ADDITIONAL INFORMATION
To reinstall the module, attach all of the cables and tubes that were removed.
