How to replace the FTN-R sample loop on an Arc HPLC Sample Manager - WKB257545
Article number: 257545
OBJECTIVE or GOAL
Replace the FTN-R sample loop because of leaking from the sample loop connection to the injector pod or the need to change from the standard 50-µL sample loop to a different size.
ENVIRONMENT
- Arc HPLC Sample Manager - FTN-R
PROCEDURE
- Power-off the Sample Manager FTN-R.
- Using two 1/4-inch wrenches, separate the existing sample loop from the needle.
- Using one 1/4-inch wrench, disconnect the existing sample loop from the injector pod (port 4).
- With the new sample loop, seat the ferrule on the male end of the sample loop and attach to port 4 of the injector pod.
- The female end of the sample loop reconnects with the needle you disconnected in step 2. (Use 1/4-inch wrenches again.)
- Power-on the sample manager after the new sample loop is fully connected into the pathway.
- If necessary, reset the Sample Manager in the ACQUITY Console to regain communications.
- In the ACQUITY Console, on the top bar, in the Configure drop-down menu, select volumes and verify that you now have the correct sample loop size, or change to the correct size.
- Perform a priming of the sample loop from the Control drop-down menu, purge the syringe with solvent for at least five cycles, and then wash the needle with solvent for at least 15 seconds.
- Verify the functionality of the system with an injection.
ADDITIONAL INFORMATION
Sample loop part numbers:
- 430002012 - 50-µL Sample Loop
- 430002011 - 100-µL Sample Loop
- 430002007 - 250-µL Sample Loop
id257545, comms, UPARBSMFTN, UPARPFTN, UPARSMFTN