Skip to main content
Waters

FTN sample syringe leak test failure - WKB257068

Article number: 257068

SYMPTOMS

  • Sample syringe leak test failure
  • No bubbles seen in syringe
  • Reproducibility is satisfactory
  • Syringe is tight and intact

ENVIRONMENT

  • ACQUITY UPLC H-Class Sample Manager - FTN
  • ACQUITY UPLC I-Class Sample Manager - FTN
  • ACQUITY ARC Sample Manager - FTN
  • ACQUITY UPLC Premier Sample Manager - FTN
  • ACQUITY ARC Premier Sample Manager - FTN 

CAUSE

Test was run in 100% acetonitrile.

FIX or WORKAROUND

  1. Change to 90:10 water:acetonitrile. 
  2. Prime the sample syringe a minimum of 60 cycles.
  3. Retest.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.