Tailing peaks with ACQUITY UPLC - WKB253843

Article number: 253843

SYMPTOMS

  • Broadening and tailing peaks
  • Fresh mobile phase, wash solvents, and column
  • System is leak-free

ENVIRONMENT

  • ACQUITY UPLC Column Heater 

CAUSE

Void caused by damaged or incorrectly positioned PEEK ferrule on the column inlet tubing

FIX or WORKAROUND

Replace the PEEK ferrule (and collet if necessary) on the column inlet tubing. Ensure that the ferrule is about 3 mm away from the outlet tube and that the collet is touching the back of the ferrule (see Figure 1 in Additional Information).

ADDITIONAL INFORMATION

Correct positioning of finger thight fitting.PNG

Figure 1: Correct positioning of finger-tight fittings

PEEK ferrule part number: 700003114

Collet: 700003115