Sample manager leak sensor error message results from leak at head of column - WKB253660
Article number: 253660
SYMPTOMS
- Sample manager leak sensor error
- Solvent observed in leak sensor above the injection valve of the Sample Manager - FTN
- Solvent leak traced back to the head of the column
ENVIRONMENT
- ACQUITY UPLC Sample Manager - FTN
- ACQUITY UPLC Column Manager - Active
CAUSE
Active preheater is not seated correctly to the head of the column.
FIX or WORKAROUND
- Disconnect the active preheater from the column.
- Verify that the ferrule does not look crushed or damaged.
- Slide the ferrule to within roughly 1 mm of the end of the tubing.
- Reseat the the active preheater to the end of the column.
- Hand-tighten snuggly but do not over-tighten.
ADDITIONAL INFORMATION
