Two Empower systems on one LAC/E32 stopped overnight - WKB25098

Article number: 25098

SYMPTOMS

  • Two LC systems on the same LAC/E32 stopped acquiring data overnight
  • "General instrument error" seen in Empower

ENVIRONMENT

  • Empower 3
  • Agilent LC
  • Alliance HPLC System

CAUSE

The InstrumentServer.exe process crashed.

FIX or WORKAROUND

  1. Reboot the LAC/E32.
  2. Restart the acquisition.

ADDITIONAL INFORMATION

id25098, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.