ACQUITY UPLC Sample Manager gives "Sample pressure too low" error - WKB250919
Article number: 250919
SYMPTOMS
- Sample Manager gives "Sample pressure too low" error during injection
ENVIRONMENT
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC I-Class Sample Manager - Fixed Loop
CAUSE
The sample syringe is not primed well or has air in it.
FIX or WORKAROUND
Prime the sample syringe for 40 cycles (do not include the wash syringes for this prime).
ADDITIONAL INFORMATION
It is recommended that the sample syringe be primed 10-20 cycles daily, before starting the instrument.