Low response on 2465 ECD - WKB250163
Article number: 250163
SYMPTOMS
- Failing standards
- Responses are noticeably lower for samples when compared with a second system using the same samples, mobile phases, flow cell, and column
- Polishing the flow cell does not correct the issue
ENVIRONMENT
- 2465 Electrochemical Detector
CAUSE
Worn flow cell
FIX or WORKAROUND
Replace the flow cell.
ADDITIONAL INFORMATION
What are the Flow Cell options for the 2465 ECD and the 3465 ECD? - wkb202136
id250163, 2465, eluent