Error during injector purge with Alliance 2695: Sample syringe over pressure - WKB25000

Article number: 25000

SYMPTOMS

  • "Sample syringe over pressure" error occurs when purging injector

ENVIRONMENT

  • Alliance HPLC system
  • 2695 Separations Module

CAUSE

Clog in injector

FIX or WORKAROUND

  1. Disconnect the Syringe-V3 tube from V3 and repeat the injector purge.
  2. Reconnect the tube.
  3. Perform injector purge and compression check to confirm that the error is resolved and that no leaks are present.

ADDITIONAL INFORMATION


id25000, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB