Error during injector purge with Alliance 2695: Sample syringe over pressure - WKB25000
Article number: 25000
SYMPTOMS
- "Sample syringe over pressure" error occurs when purging injector
ENVIRONMENT
- Alliance HPLC system
- 2695 Separations Module
CAUSE
Clog in injector
FIX or WORKAROUND
- Disconnect the Syringe-V3 tube from V3 and repeat the injector purge.
- Reconnect the tube.
- Perform injector purge and compression check to confirm that the error is resolved and that no leaks are present.
ADDITIONAL INFORMATION
id25000, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB