Error during injector purge with Alliance 2695: Sample syringe over pressure - WKB25000
Article number: 25000
SYMPTOMS
- "Sample syringe over pressure" error occurs when purging injector
ENVIRONMENT
- Alliance HPLC system
- 2695 Separations Module
CAUSE
Clog in injector
FIX or WORKAROUND
- Disconnect the Syringe-V3 tube from V3 and repeat the injector purge.
- Reconnect the tube.
- Perform injector purge and compression check to confirm that the error is resolved and that no leaks are present.