Agilent LC pump overpressurizes (bad active inlet valve cartridge) - WKB249781
Article number: 249781
SYMPTOMS
- Agilent LC pump overpressurizes
- Agilent LC pump shows random drops in pressure
- Manual flow rate accuracy test failure
- Leak test failure
ENVIRONMENT
- Agilent LC
CAUSE
Bad active inlet valve cartridge
FIX or WORKAROUND
Replace the active inlet check valve assembly (5062-8562).
ADDITIONAL INFORMATION
Find directions on how to replace the active inlet check valve in Repairing/Replacing the Pump Head Assembly, pages 5-7.
This part can be used on models G1310A, G1311A/C, G1312A/C, G1376A, and G2226A for a maximum pressure of 400 bar.
(For models G1310B, G1311B, G1312B, G7111A/B, and G7112B, use part G1312-60020 for a maximum pressure of 600 bar.)
id249781, checkvalve