SYMPTOMS

  • Leak from bottom of the instrument

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Fitting for solvent line is loose.

FIX or WORKAROUND

Tighten the loose fitting. 

ADDITIONAL INFORMATION

 