Skip to main content
Waters

Solvent leaks from bottom of Alliance HPLC System - WKB248991

Article number: 248991

SYMPTOMS

  • Leak from bottom of the instrument

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Fitting for solvent line is loose.

FIX or WORKAROUND

Tighten the loose fitting. 

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.