ACQUITY UPLC H-Class QSM fails system leak test - WKB247193

Article number: 247193

SYMPTOMS

  • System leak test failure
  • Leak test successful on both Primary and Accumulator pump heads

ENVIRONMENT

  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

CAUSE

Leaking vent valve P2- inlet filter tubing 

FIX or WORKAROUND

  1. Replace the vent valve P2- inlet filter tubing (430002319).
  2. Verify that system now passes leak test.

ADDITIONAL INFORMATION