Skip to main content
Waters

ACQUITY PDA or TUV pops tubing at flow cell inlet or post-column - WKB245305

Article number: 245305

SYMPTOMS

  • Tubing keeps popping off, either at the flow cell inlet or post-column 
  • No flow from outlet of PDa when a low flow (0.1 mL/min) is started, but tubing still pops, accompanied by a pressure increase of 3000 psi when flow cell is connected
  • Ensure new tubing does not resolve issue per wkb93638

ENVIRONMENT

  • ACQUITY PDA
  • ACQUITY TUV

CAUSE

Blocked flow cell

FIX or WORKAROUND

  1. Flush the flow cell.  
  2. If flushing doesn't resolve the issue, replace the flow cell. 

  PDA

  • Analytical Flow Cell (10-mm) (205015017) Standard 
  • High-sensitivity Flow Cell (25-mm) (205015019)
  • High-throughput Flow Cell (3-mm) (205015006)
  • ACQUITY PDA 5 mm SS FC (205000612)

TUV

  • ACQUITY UPLC 10 mm Cell TUV, Analytical (205015016) Standard 
  • ACQUITY Cell TUV, Standard (205015000)
  • ACQUITY UPLC TUV Flow Cell, Low Flow (205015001)
  • ACQUITY UPLC 25 mm Cell TUV, HS FC (205015018)
  • nanoTUV Flow Cell (205015013)
  • ACQUITY UPLC TUV 5 mm SS FC (205000610)
  • ACQUITY UPLC TUV 5 mm Titanium FC (205000611)

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.