Pressure fluctuations observed on ACQUITY Arc Premier solvent manager - WKB244971

Article number: 244971

SYMPTOMS

  • Pressure fluctuations observed on ACQUITY Arc Premier solvent manager

ENVIRONMENT

  • ACQUITY Arc Premier
  • ACQUITY Premier

CAUSE

Failing check valves on both primary and accumulator pump heads

FIX or WORKAROUND

Replace the check valves:

700005414 - CARTRIDGE, i2V, BIOACQUITY

700005415 - CHECK VALVE, ACCUMULATOR, TI

ADDITIONAL INFORMATION