For ACQUITY UPLC, what is the degasser health check? - WKB243938
Article number: 243938
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC BSM
- ACQUITY Arc QSM
- ACQUITY Arc Bio QSM
ANSWER
When a degasser hardware failure is indicated, as an initial troubleshooting step, the health check attempts to clear any excessive condensate from the degasser vent line, which can trigger a false alarm. The health check can be configured to prime all the lines as well.
ADDITIONAL INFORMATION
id243938, A-30SDM, UPAQSM, UPBINARY, UPBSM+, UPBSMIVD, UPIBSM, UPIBSM+, UPIBSMIVD, UPLCCOLHTR, UPNBINARY, UPPBSM, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR