ACQUITY UPLC Dynamic Leak test fails: "Primary Pump unable to compress" - WKB243889
Article number: 243889
SYMPTOMS
- Pump able to prime with good flow
- When running Dynamic Leak test, the Primary starts to compress but fails to reach the set pressure (error code: unable to compress)
- When running Dynamic Leak test, the Primary fails to maintain the target pressure until the end of the Dynamic Leak test (error code: unable to maintain pressure)
- Check valves are new for both Accumulator and Primary pump heads
ENVIRONMENT
- ACQUITY UPLC Binary Solvent Manager (BSM)
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC H-Class BSM
- ACQUITY UPLC I-Class BSM
CAUSE
Worn seals, rebuild pump head
FIX or WORKAROUND
- Replace high pressure seal in pump head.
For ACQUITY Classic and H-class, part number is 700004747
For ACQUITY I-class, part number is 700010583