System precision fails to meet height specification - WKB242283

Article number: 242283

SYMPTOMS

  • System precision fails to meet height specification
  • New mobile phase and column do not resolve issue
  • Passes specification when tested with pre-mixed mobile phase
  • Replacing the GPV does not resolve

ENVIRONMENT

  • ACQUITY Arc
  • Arc HPLC System

CAUSE

Leak at the primary inlet check valve filter

FIX or WORKAROUND

Replace the in-line filter (700002912 - ASSY, FILTER, IN-LINE, SS FRIT).

ADDITIONAL INFORMATION

 

id242283, checkvalve, eluent, gradient proportioning valve, UPARQSM