Low sensitivity or blank chromatograms but signal as expected when using direct infusion - WKB240722

Article number: 240722

SYMPTOMS

  • Low sensitivity or blank chromatograms but signal as expected when using direct infusion
  • Passes sample syringe leak test
  • Fails needle seal readiness test

ENVIRONMENT

  • MassLynx
  • Empower
  • UNIFI
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • Quadrupole MS

  • Tof MS

CAUSE

The needle seal is worn and the needle no longer makes a good seal.

FIX or WORKAROUND

  1. Characterize needle seal.
  2. Run the needle seal readiness test.

ADDITIONAL INFORMATION

 