Skip to main content
Waters

Broken fitting on the ACQUITY UPLC Column Heater's active preheater (205001774) - WKB240457

Article number: 240457

SYMPTOMS

  • Column trough failure on position 1 or 2

ENVIRONMENT

  • ACQUITY UPLC H-Class PLUS

CAUSE

Broken APH retaining clips. APH not secure in APH port.

FIX or WORKAROUND

Order a new active preheater Part Number (205001774).

ADDITIONAL INFORMATION

Engineering code 279003770 relates to Part Number 205001774 (ACQ UPLC APH SS 12.5LG - PLUS).

Not able to find a solution? Click here to request help.