Poor area reproducibility on Arc HPLC - WKB240418
Article number: 240418
SYMPTOMS
- Sample areas/height inconsistent
- Test shows the sample amount drawn by injector varies
ENVIRONMENT
- Arc HPLC
CAUSE
Air in the syringe
FIX or WORKAROUND
- Prime the syringe.
- Clear any air bubbles by tapping on the syringe barrel on the upstroke during priming.
- If possible, perform the sample syringe leak test to demonstrate sampling components are leak-free.
ADDITIONAL INFORMATION
id240418, A-30SM, UPISMFTN, UPPFTN, UPSMFTN, UPSMFTN+