High area RSD on ACQUITY UPLC - WKB240182
Article number: 240182
SYMPTOMS
- Poor peak area %RSD for ACQUITY UPLC (1.5%–2.0% from three replicate injections)
- Sample Manager passes all diagnostic tests, including the sample syringe leak test
- Needle and loop volume characterization passes with the same measured volumes each time
ENVIRONMENT
- ACQUITY UPLC
CAUSE
Leak in the BSM; fails dynamic leak test for one or both pumps.
FIX or WORKAROUND
Rebuild the leaking pump head. Replace the seals and check valve. Replace the pump head and seal wash housing if necessary.