Alliance column heater does not reach set temperature - WKB238680

Article number: 238680

SYMPTOMS

  • 2695 column heater is not heating
  • LEDs on column heater PCB operate properly
  • Fan spins properly

ENVIRONMENT

  • 2695 Separations Module
  • Alliance Series Column Heater

CAUSE

Worn heating coil  

FIX or WORKAROUND

Replace the heating coil (WAT270685).

ADDITIONAL INFORMATION

 