Skip to main content
Waters

Alliance column heater does not reach set temperature - WKB238680

Article number: 238680

SYMPTOMS

  • 2695 column heater is not heating
  • LEDs on column heater PCB operate properly
  • Fan spins properly

ENVIRONMENT

  • 2695 Separations Module
  • Alliance Series Column Heater

CAUSE

Worn heating coil  

FIX or WORKAROUND

Replace the heating coil (WAT270685).

ADDITIONAL INFORMATION

 

id238680, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB