Skip to main content
Waters

No communication with Ethernet components in ChromScope - WKB237687

Article number: 237687

SYMPTOMS

  • Alarm for 2545 and 2489 in ChromScope Run window
  • Will not reset
  • Will not re-populate DHCP after clearing and reboot of modules

ENVIRONMENT

  • ChromScope
  • Prep 150 LC System
  • 2545 Quaternary Gradient Pump
  • 2489 Dual-Wavelength Absorbance Detector

CAUSE

Unknown

FIX or WORKAROUND

  1. Power-down all Ethernet modules.
  2. Close out software and shut down PC.
  3. Power-on modules and allow them to reach idle state.
  4. Power-on PC and log in to Windows.
  5. Re-launch ChromScope and check for alarms.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.