Low signal from injections with ACQUITY SM-FTN - WKB237523
Article number: 237523
SYMPTOMS
- Much lower than expected signal from injected samples
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
The needle and seal holder tubing are connected to the wrong ports on the injection valve, so no sample is being drawn up by the needle.
FIX or WORKAROUND
Ensure that the seal holder tube is connected to port 1 and the FTN needle is connected to port 4 of the injection valve.
ADDITIONAL INFORMATION
id237523, A-30SM, UPISMFTN, UPPFTN, UPSMFTN, UPSMFTN+