ANSWER

Basically, the order is as follows.

■ H-Class
When using 30cm CH:
30cm CH → QSM/FTN in no order → PDA
When not using 30cm CH:
QSM/FTN/CM in no order → PDA

■ Original UPLC
When using 30cm CH:
30cm CH→SM→BSM→PDA
When not using 30cm CH:
Those with built-in SM or CM hub → BSM → PDA

*In any system, first determinbe whether the built-in switch (hub) or an external switch is used.
Check the back of the instrument; if it has multiple LAN ports, it means that the built-in switch is used.
If using the built-in switch, power-on from that module to ensure recognition of other modules.
* 30cm CH is an accessory of the SM, so it starts before the SM. SO starts after SM.

ADDITIONAL INFORMATION

There is no particular order when powering-off.