Alliance 2695 plunger homing fault (0) - WKB235241
Article number: 235241
SYMPTOMS
- Plunger Homing Fault (0) error on 2695
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
The check valves are stuck.
FIX or WORKAROUND
- Replace the check valves (part 700000254).
- Reboot the separations module with the vent valve open.
- Close the vent valve after initialization.
- Prime and run instrument as intended.