Alliance 2695 plunger homing fault (0) - WKB235241

Article number: 235241

SYMPTOMS

  • Plunger Homing Fault (0) error on 2695

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

The check valves are stuck.

FIX or WORKAROUND

  1. Replace the check valves (part 700000254).
  2. Reboot the separations module with the vent valve open.
  3. Close the vent valve after initialization.
  4. Prime and run instrument as intended.

ADDITIONAL INFORMATION