Dark current error and grating failure on ACQUITY TUV - WKB23222
Article number: 23222
SYMPTOMS
- Dark current error
- Grating failure
ENVIRONMENT
- ACQUITY TUV
CAUSE
Faulty lamp
FIX or WORKAROUND
Replace the lamp (201000273).
ADDITIONAL INFORMATION
id23222, A-30UV, UPARPTUV, UPPTUV, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC