Rotary tray move h/w error on ACQUITY SM-FL when running from ACQUITY SO under SCIEX OS control - WKB231506

Article number: 231506

SYMPTOMS

  • Intermittent rotary tray move h/w errors when running from ACQUITY SO
  • It may run 10-20 injections successfully before error
  • Plates are configured properly
  • Happens with well plates or vial trays
  • All SM calibrations were done successfully
  • No errors when running from tray 1 in SM
  • No issues when running under MassLynx control

ENVIRONMENT

  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC SM-FL
  • ACQUITY UPLC SM-FTN
  • ACQUITY UPLC SO
  • SCIEX OS, multiple versions
  • DP2019R2
  • SCIEX OS Support Layer v1.1

CAUSE

Defect in Waters DP2019R2/SCIEX OS Support Layer v1.1. The issue is resolved in DP2022R1/SCIEX OS Support Layer v1.2.

FIX or WORKAROUND

  1. Upgrade to DP2022R1 and SCIEX OS Support Layer v 1.2 (see WKB125305 for details).
  2. Verify that instruments run properly without error.

ADDITIONAL INFORMATION

 

id231506, A-30SM, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPORGANIZE, UPORGIVD, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO